General Specifications |
Function |
Direct UV light absorption measurement for COD, BOD (by correlation of COD), TSS, pH, Temperature |
Model |
P7111 |
Installation |
Wall mounted or Freestanding |
Protection Class |
IP65 |
Enclosure |
• Dimension: H1250* W550* D430mm • Weight: 20Kgs • Material: Steel |
Range |
• Chemical Oxygen Demand: 0~200mg, or as per user selective range • Biochemical Oxygen Demand: 0~200 mg/L • Total Suspended Solids: 0~200 mg/L • pH: 0-14 |
Accuracy |
F.S <±5 % |
Repeatability |
<±5 % |
Controller Display |
7” LCD color touch panel display |
Response Time |
< 60s |
Calibration |
Automatic calibration. Can switch to manual calibration, high- and low- point |
Self-Diagnosis |
The system is continuously self-diagnosis. The error management helps to identify problems. (Display errors and save historical data) |
Operating temperature |
0~ 50°C |
Operating humidity |
< 95% RH |
Sample conditions |
• Sample temperature: 0 to 45 °C • Sample flow: Continuous, 50 to 1,000 ml/min • Suspended solids: 200 microns or less |
Communication Outputs |
• Analog: 4 – 20 mA isolated, RS-485(Modbus) • Relay: 1 x alarm, 1 x system error • Digital: RS 485/ Modbus |
Electrical Conditions |
Power supply ranges:115 / 230 VAC, 50 / 60 Hz,<500 Watt |
Language Support |
Chinese, English |
Environmental Conditions |
• Ambient operating temperature: 0 – 60 °C • Ambient operating humidity: Up to 95 % RH non-condensing |